uRADMonitor
KeymasterThanks guys!
uRADMonitor
KeymasterIt looks like it lost connection. What is the firmware version on your unit?
uRADMonitor
KeymasterDefinitely, see this: https://www.uradmonitor.com/topic/extend-the-maps/
I’ve been a little caught with the hardware work for model D and A3, but this will be done as soon as possible, part of the new frontend version.
uRADMonitor
Keymaster0.005 for your tube seems very reasonable, however you would need to test it against a calibrated dosimeter.
I might do that as well, should I find 2x SI16BG tubes. Do you know any good source for them?
uRADMonitor
KeymasterMost likely yes. Will get fixed, it’s just a very busy period atm.
uRADMonitor
KeymasterFirst keep the value in the schematics. Next try a bigger value like 100pF. Hope this helps, keep me posted.
uRADMonitor
KeymasterLatest firmware is v123 . Here is what you need: https://www.uradmonitor.com/firmware-upgrade-guide/
uRADMonitor
KeymasterHello Sandra, please order it on the Products page: https://www.uradmonitor.com/products/ . To get it before Christmas you should also add the “Express shipping” option to your chart.
uRADMonitor
KeymasterI went to the DB to do the changes manually, then I save the 1100000A is already assigned to your user id. Please click the Dashboard and post a screenshot with what you see there: https://www.uradmonitor.com/dashboard/
Thanks!
uRADMonitor
KeymasterHello Maarten,
Thanks for these observations. I added ppm for co2 and ch2o, this refers to the concentration of these gases in air. Please check the various thresholds and their effects on human health.
For VOC we have no standard unit, it is more of a score computed out of REDOX chemical reactions on the sensor’s crystal. This will get improved by adding a new adaptive algorithm that analyses few minutes of previous data. Bosch already has this, and we might use their solution directly.
Thank you,
RaduuRADMonitor
KeymasterThe unit will appear automatically listed in the Dashboard, IF you are in the same LAN as your unit. We’re using IP address to do the matching.
If this is not possible let me know, and I will change it manually.
Thanks!
uRADMonitor
KeymasterThanks for the fix, no compiler option but a bug originally reported by Akos. Thanks for reporting this.
uRADMonitor
KeymasterThis oscillation is not normal.
I looked over the data and can advise you the following:
https://raw.githubusercontent.com/radhoo/uradmonitor_kit1/master/pcb/1.2.105/circuit/kit1_105_sch.png1. change the compensation capacitor (C9)
2. change the geiger tubeuRADMonitor
KeymasterIt looks like a tube failure.
You can either ship it back for a replacement, or change the tube directly as it only requires simple soldering.
uRADMonitor
KeymasterFlag changed. Thanks!
-