Home Forum Hardware New A3 – Temperature readings too high

Viewing 3 posts - 1 through 3 (of 3 total)
  • Author
    Posts
  • #6086
    Radu
    Keymaster

    Do you have more data on this? More correlation sets to other instruments? There is an offset that can be adjusted here, and I’d like to do that for your unit.

    #6269
    WeakNerve
    Participant

    It seems to be a much bigger difference when it is colder. Here is a quickly pasted together comparison to a nearby (1km) weather station, the last week of October. Higher temperature being the A3.

    I looked up a warmer day in September.
    16th September 02:00 Weather station reports +11.2c while the A3 reports +14.1c.
    At 16:00 the same day Weather station reports +13.1c while the A3 reports +15.7c.

    So putting in an offset will get it in the right direction, but to me it seems very difficult to get it to be correct all year around.

    #6277
    Radu
    Keymaster

    Temperature compensation offset changed to -2.4 . Let’s see how this goes.